Keywords
Cardiac resynchronisation therapy, conduction system pacing, leadless pacemaker, left bundle area pacing, leadless pacemaker, subcutaneous ICD
Take-home messages
- The pacing site has a relevant impact on cardiac function.
- Conduction system pacing offers more physiological ventricular activation, a better haemodynamic performance than right ventricular pacing, and is equal or even superior to biventricular pacing.
- Leadless pacemakers and subcutaneous ICDs have overcome complications presented by transvenous leads and device pockets.
- Leadless pacemakers and subcutaneous ICDs still have limitations in terms of pacing site and antitachycardia pacing.
- Clinicians and electrophysiologists must carefully choose the most appropriate device and pacing site, evaluating the needs and risks of each patient.
Abbreviations
CRT: cardiac resynchronisation therapy
CSP: conduction system pacing
HBP: His bundle pacing
HOT-CRT: His bundle-optimised CRT
LBBAP: left bundle branch area pacing
LBBP: left bundle branch pacing
LOT-CRT: left bundle-optimised CRT
sICD: subcutaneous implantable cardiac defibrillator
Introduction
Since the first totally implantable pacemaker in 1958, cardiac pacing from the right ventricle (RV) apex has been lifesaving for patients with complete atrioventricular block. The first pacemakers had a battery lasting only 8 hours and were fixed-rate. Battery life reached 11 months in 1960, and the introduction of lithium-iodine batteries allowed a significant longevity increase of up to 10 years.
Adaptive rate and dual chamber pacing improved exercise performances, particularly in sinus node disfunction. But apical RV pacing creates a left bundle branch (LBB) block morphology with a delayed left ventricular (LV) activation. Substantial ventricular pacing (>20-40%) was found to increase the risk of pacing-induced cardiomyopathy, heart failure (HF) hospitalisation, and atrial fibrillation (AF) even when atrioventricular (AV) synchrony was preserved [1]. To minimise unnecessary ventricular pacing, several algorithms have been developed to promote intrinsic atrioventricular conduction, extending AV intervals in AAI mode, with an automatic switch to DDD mode if no ventricular rhythm can be found.
These negative consequences underlined how much the pacing site contributes to cardiac output and prompted the search for additional physiological pacing sites. However, RV high septal pacing showed no significant differences in HF hospitalisation, mortality, AF burden, or plasma B-type natriuretic peptide levels when compared with apical pacing.
Central Illustration. Timeline of the evolution of CIEDs.
ATP: antitachycardia pacing; CRTD: cardiac resynchronisation therapy device; HOT: His optimised therapy; LBB: left bundle branch; LOT: left bundle optimised therapy; RV: right ventricle; RVOT: right ventricle outflow tract; RSP: right septal pacing
Cardiac resynchronisation therapy
In 1994, Cazeau [2] demonstrated that biventricular (BiV) pacing with a lead on the LV free wall can reduce QRS duration and mitral valve regurgitation in HF patients with LBB block (LBBB). Cardiac resynchronisation therapy (CRT) can decrease HF hospitalisations and cardiac death, but up to 30% of patients are non-responders.
To reduce the number of CRT non-responders, echo-guided or automatic algorithm-based parameter optimisation can detect AV and interventricular intervals, leading to a better haemodynamic response than default programming. LV intracardiac impedance measurements via the RV coil, the RV tip for current injection, and LV-CS proximal and distal electrodes for voltage measurement, can detect LV volume changes. In idiopathic and ischaemic congestive HF patients, a strong correlation between changes in intracardiac impedance and LV stroke volume in acute tests at the CRT implant has been found [3-4]. This good correlation is limited to acute settings because the tilt angle between RV and LV leads can influence results in chronic implants.
Some other, smaller studies have demonstrated that in acute and chronic settings, bifocal RV pacing is more effective than apical RV pacing, and is sometimes not inferior to biventricular pacing, for the improvement of haemodynamics. Triangle ventricular pacing was superior to biventricular pacing in terms of dP/dt and cardiac output. Both triangle pacing modes (two right plus one left, or one right plus two left leads) improved the distance covered during the 6-minute walking test, the quality-of-life score, the LV end-diastolic and end-systolic volumes, and the ejection fraction, more than conventional biventricular pacing.
In 2012, multipoint LV pacing was introduced. This pacing mode improved acute haemodynamic response assessed by pressure-volume loops [5] better than single-site pacing. The rationale of multipoint pacing is based on the fact that stimulation with a single electrode (point stimulation) produces a convex excitation front, whereas stimulation with a line of electrodes (line stimulation) produces an almost flat excitation front. Multipoint pacing captures a larger area of myocardium and improves transventricular activation time. Bifocal RV pacing plus multipoint LV pacing could theoretically be more effective than conventional biventricular pacing and triangle pacing too, as demonstrated in one of our patients, an 81-year-old man with ischaemic cardiomyopathy with permanent AF. Connecting the RV apical lead to the atrial port of a multipoint CRT-D was possible to obtain a quadrifocal stimulation. At hospital discharge, all triangle pacing configurations showed an aortic velocity time integral increase in the order of 10% and, in quadrifocal pacing, 22% as compared to spontaneous rhythm. Left ventricular ejection fraction (LVEF) increased from 25% at baseline to 42% at 1-year follow-up.
In 2000, Deshmukh [6] proposed His bundle pacing (HBP) to preserve normal electrical activation of the ventricles and to prevent ventricular dissynchrony. But His bundle pacing is challenging because the target area is anatomically limited and pacing thresholds are higher than for septal and apical areas at implantation and in 10-14% of cases can increase over 3 months. Despite the fact that approximately two-thirds of patients have non-selective HBP at programmed output, it is not significantly different from selective HBP, and total effectiveness is 84%.
In 2017, Huang [7] first reported left bundle branch pacing (LBBP) for a patient with LBBB. A dedicated lead was deployed deep in the left interventricular septum to capture the LBB beyond the block site. LBBP produced early activation of the LV with a right bundle branch block pattern in lead V1. Also, LBBP can be difficult to achieve, due to the limited width of the LBB, and sometimes pacing is done at the left fascicular or left ventricular septal site: these sites are considered as part of the LBB.
Vijayaraman [8] prospectively investigated the feasibility and safety of LBB area pacing (LBBAP). The success rate of permanent LBBAP reached 93% in patients with a paced QRS duration of 136±17 ms. The low capture threshold and high ventricular amplitude remained stable at the median follow-up of 3 months. Compared with RV pacing, LBBAP improved the primary composite outcome of all-cause mortality, HF hospitalisation, or upgrade to BiV pacing. Thus, LBBAP is an alternative pacing form that is able to overcome the limitations of HBP. However, there are some advantages and disadvantages with both of these areas for conduction system pacing (CSP): HBP may be preferred in patients with tricuspid valve dysfunction, or prosthesis or transcatheter repair, while LBBAP may be preferred in patients with scheduled AV node ablation, infranodal block, previous or scheduled transcatheter aortic valve implantation (TAVI), or aortic valve surgery.
Conduction system pacing for heart failure
In some small clinical trials, both HBP and LBBAP were non-inferior or, in some cases, superior to BiV-CRT in terms of improved New York Heart Association class and LVEF, with no significant differences in HF hospitalisation and mortality over limited follow-up time.
A suboptimal resynchronisation with CSP is still present in some cases, as in BiV-CRT: to optimise the quality of cardiac resynchronisation, His bundle-optimised CRT (HOT-CRT) [9] and left bundle branch-optimised CRT (LOT-CRT) [10] were proposed, adding a coronary sinus lead to a CSP. Both HOT and LOT incremented electrical resynchronisation compared with HBP, LBBAP, and BiV-CRT.
Remote monitoring
In 2000, Home Monitoring was introduced by Biotronik to allow daily remote control of heart rhythm and device function. Several other companies introduced iterations of remote monitoring (RM), and now RM is the standard of care for the early detection of leads malfunction, battery depletion, and new onset of atrial and ventricular arrhythmias. Daily alert transmissions can facilitate the early detection and quantification of AF episodes and arrhythmia burdens that may prompt clinical reactions, preventing adverse events such as stroke, inappropriate shock therapy, and HF. RM reduces the number of health care visits and is cost-effective. It also reduces patient costs, reducing the need to travel to in-person visits and take time off from work for patients and caregivers both. RM confirmed its utility particularly during the COVID19 pandemic when lockdown forced patients to stay home.
Algorithms to improve implantable device performance
Over the past two decades, cardiac implantable electronic devices (CIEDs) have progressively incorporated diagnostic algorithms aimed at the early detection of HF decompensation.
The earliest algorithm, OptiVol (2004-2005; Medtronic, Galway, Ireland), was based on intrathoracic impedance, a surrogate marker of pulmonary congestion. Intrathoracic impedance is measured by delivering a low-amplitude current between the intracardiac lead and the device can. A decrease in impedance reflects increased thoracic fluid content and was shown in early studies to precede HF events by approximately 10-15 days [11]. Subsequent trials such as SENSE-HF demonstrated limited sensitivity and predictive accuracy of single-parameter monitoring.
Therefore, device diagnostics evolved toward the integration of additional physiological signals, such as heart rate, atrial fibrillation burden, patient activity, and biventricular pacing percentage (as in Medtronic’s Cardiac Compass device interrogation reports). The PARTNERS-HF study [12] demonstrated that combining multiple device-derived parameters – including impedance, nocturnal heart rate, arrhythmia burden, and activity – significantly improved prediction of HF hospitalisations compared with single metrics.
Subsequently, fully integrated predictive algorithms such as TriageHF implemented probabilistic models combining multiple diagnostics to stratify short-term HF risk. These systems marked the transition from descriptive monitoring to automated risk prediction based on temporal trends across multiple physiological domains.
The most advanced stage of this evolution is represented by multisensor algorithms (Heart Logic by Boston Scientific), reflecting different aspects of HF pathophysiology, including heart sounds (S1 and S3), thoracic impedance, respiratory rate, nocturnal heart rate, and physical activity. The algorithm generates a composite daily index, validated in the MultiSENSE trial [13], which demonstrated approximately 70% sensitivity and a median lead time of over one month before HF events. Subsequent studies such as MANAGE-HF confirmed the clinical utility of this approach in guiding early therapeutic interventions.
Also, the HeartInsight algorithm (Biotronik) integrates multiple physiological parameters, including heart rate metrics, premature ventricular contractions, atrial burden, heart rate variability, patient activity, and thoracic impedance, capturing autonomic, electrical, and haemodynamic aspects of HF. The SELENE-HF registry confirmed its ability to identify periods of increased HF risk in real-world populations.
This progression from a single-parameter model to integrated multi-sensor platforms reflects both technological advancements and a deeper understanding of HF pathophysiology, where congestion, autonomic imbalance, arrhythmias, and reduced functional capacity interact dynamically. These device-derived parameters are transmitted daily through RM systems to the hospital, allowing early clinical evaluation, timely patient contact, and prompt optimisation of heart failure therapy before overt decompensation occurs.
Atrial pacing
Interatrial conduction delay can increase the risk of atrial arrhythmias. In 1999, in the PIPAF protocol, the dual site atrial pacing (right auricula + coronary sinus ostium), despite a positive trend, was not statistically effective at preventing atrial fibrillation. More recently, Bachmann’s bundle pacing proved effective at significantly decreasing atrial arrhythmia burdens, recurrence, and de novo incidence compared with right atrial septal pacing and right atrial appendage pacing [14].
Leadless pacemakers
Complications are reported in up to 10-15% of cases of pacemaker implants, including venous thrombosis and obstruction of superior caval system branches, significant regurgitation of the tricuspid valve, infections, chronic lead failure due to insulation problems, and lead fracture. Some of them, such as haematomas, skin erosion, or life-threatening pocket infections, require the extraction of the whole pacemaker system, including the leads. To eliminate pacemaker leads, the Nanostim leadless pacing system (Abbott, Abbott Park, IL, USA) and the Micra Transcatheter pacing system (Medtronic, Galway, Ireland) were introduced in 2012 and 2013, respectively. Both systems are totally implantable in the RV, using an 18 or 24 F femoral approach. Both systems are theoretically retrievable. In a retrospective analysis, when compared to traditional RV pacing, leadless pacemakers (LP) showed a lower incidence of pacing-induced cardiomyopathy, but this result was not confirmed in the other two studies.
In 2020, an AV-synchronised pacing device was approved: based on the three-axis accelerometer algorithm, which allows the sensing of atrial mechanical activity, the device is capable of VDD pacing. In 2022, a true dual chamber LP with two distinct devices was introduced to pace the right atrium and RV in a synchronised mode.
An internal jugular vein approach has been suggested to overcome the problem of the tortuous femoral vein, taking advantage of a shorter route and making the delivery system easier to manipulate. This superior access results in a more natural curve towards the RV septum.
LPs for conduction system pacing are under investigation to overcome the limitations of the sites that can actually be paced (RV pacing is inferior to CSP for avoiding pacing-induced cardiomyopathy).
Twelve-lead ECG and pacing site
QRS morphology in standard 12-lead electrocardiography (ECG) can indicate the ventricular zone that is paced (Figure 1):
- The RV apex shows an LBBB with superior axis.
- The mid-basal interventricular septum shows an LBBB with horizontal axis.
- Epicardial LV pacing in a coronary sinus branch shows a large right branch bundle block.
Biventricular pacing can show shorter QRS with variable axis and morphology depending on the site (whether it is more posterior or lateral, apical or basal).
LBB capture can be recognised by: (1) the existence of LBB potential and the interval from LBB potential to R-wave peak in lead V6 equal to the stimulus to R wave peak time (RWPT) in V6; (2) transition from non-selective to selective LBB capture with decrementing output (or with programmed stimulation); (3) retrograde activation of the HB electrogram; (4) abrupt shortening and then short and constant stimulus to RWPT in V6 during lead placement; (5) a longer V6-V1 interpeak interval defined by RWPT in V6 and V1; or (6) a shorter stimulus to RWPT.
Figure 1. Variation in ECG readings depending on pacemaker placement
From A to J: QRS morphologies with pacing at different sites in the same patient. L: LBB pacing in a different patient.
IVS: interventricular septum; LVd: left ventricle distal; LVp: left ventricle proximal; RVA: right ventricle apex
Implantable cardiac defibrillators
In February 1980, the first defibrillator – with epicardial leads and the device weighing 300 grams – was implanted in an abdominal pocket. Some years later, transvenous leads and a weight reduced to 90 grams allowed implants to be placed in a subpectoral pocket. Another fundamental evolution was the introduction of antitachycardia pacing (ATP) to treat ventricular tachycardias without delivering a shock. Patients with indications for implantable cardiac defibrillators (ICD) often have a reduced ejection fraction and an LBBB. Therefore, in August 1998, we performed the first biventricular ICD using a Y adapter because devices with a dedicated LV lead port were not available at that time [15].
The same risks of infection and lead dislodgement that are present in pacemakers also appear in ICDs. In 2008, the first subcutaneous ICD (sICD) was introduced [16], using a parasternal electrode and a left lateral thoracic pulse generator. This configuration was as effective as a transvenous ICD for terminating induced ventricular fibrillation, albeit with a significantly higher mean (±standard deviation) energy requirement (36.6±19.8 J vs. 11.1±8.5 J). In the EFFORTLESS Study, 985 patients had 16.9% inappropriate shocks in a 5-year follow-up, due to T wave oversensing in 68.3% of cases and, in 18.7%, to supraventricular tachycardias or atrial fibrillation. In 2015, an algorithm using morphological characteristics of the QRS was introduced and allowed a 39.8±11.4% reduction of inappropriate shocks. The subsequent SMART PASS filter, introduced in the second and third generations of sICDs, reduced the annual inappropriate shock rate to 3.1% and to 2.4% in the EMBLEM devices.
The two most important limitations of sICDs are pacing capability, limited to 30 seconds of antibrady period after a successful shock, and the impossibility of providing antitachycardia pacing.
To address these limitations, two different strategies were introduced: a leadless intracardiac pacemaker communicating with an sICD and a substernal extravascular ICD.
In the MODULAR ATP study, a communication was created between an sICD and a dedicated leadless pacemaker implanted in the right ventricle that was able to deliver ATP and to pace as a normal antibradycardia device.
A substernal extravascular lead was introduced in 2015 [17]. The lead has an epsilon shape with 2 pacing/sensing electrodes and 2 defibrillation coil segments; it is inserted under the sternum from the xiphoid and the left costal margin to the upper border of the heart silhouette. The device is placed either over the serratus muscle or between the serratus and the latissimus dorsi. The lead position allows pacing with 5.4±2.2 V at 8 ms width and defibrillation with 15-30 J. Unfortunately, most of the first group of implanted patients reported awareness of pacing, which they described as mild or moderate.
Conclusions/Impact on practice statement
Recent innovations in cardiac pacing are reshaping clinical practice by promoting more physiological atrial and ventricular activation, reducing pacing-induced cardiomyopathy, and improving heart failure outcomes. Conduction system pacing, particularly His bundle pacing and left bundle branch area pacing, is emerging as a valuable alternative to conventional right ventricular pacing and, in selected patients, even to traditional cardiac resynchronisation therapy. Multi-sensor remote monitoring algorithms now enable earlier detection of heart failure decompensation and more proactive patient management. Leadless technologies and modular defibrillation systems further reduce device-related complications and expand therapeutic possibilities. Overall, modern cardiac implantable devices are evolving from simple rhythm correction tools toward integrated platforms for personalised cardiovascular care. Clinicians and electrophysiologists must carefully choose the most appropriate device and pacing site, evaluating the needs and risks of each patient.