Keywords
Cardiac implantable electronic device; device malfunction; implantable cardioverter-defibrillator; pacemaker; troubleshooting
Abbreviations
CIED: cardiac implantable electronic device
CRT: cardiac resynchronisation therapy
ECG: electrocardiogram
EP: electrophysiology
HF: heart failure
ICD: implantable cardioverter-defibrillator
Take-home messages
- Most clinically relevant CIED malfunctions can be suspected before device interrogation using clinical history, physical examination, ECG, and chest radiography.
- Symptoms are often the first indication of device malfunction and should always be interpreted in relation to the patient's indication of pacing and pacing dependency.
- Not every unexpected ECG finding represents device malfunction; modern pacing algorithms frequently mimic abnormal device behaviour.
- A normal chest X-ray does not exclude electrical lead dysfunction, while an abnormal ECG almost always warrants device interrogation.
- The role of the general cardiologist is to recognise abnormalities requiring urgent EP assessment while avoiding unnecessary referrals for normal device function.
Introduction
The number of patients living with cardiac implantable electronic devices (CIEDs) continues to increase worldwide due to expanding indications for permanent pacing, cardiac resynchronisation therapy (CRT), and implantable cardioverter-defibrillator (ICD) therapy. Consequently, physicians outside specialised electrophysiology (EP) centres increasingly encounter patients with implanted devices presenting with different symptoms or incidental electrocardiographic (ECG) abnormalities.
Although device interrogation remains the cornerstone for diagnosing CIED malfunction, it is neither immediately available in many clinical settings nor necessary in every patient. Importantly, the majority of clinically significant device problems produce characteristic abnormalities that can already be suspected using a focused medical history, careful physical examination, a standard 12-lead ECG, and chest radiography.
Central illustration. Clinical troubleshooting of cardiac implantable electronic devices: a practical bedside approach for the general cardiologist.
AF: atrial fibrillation; CIED: cardiac implantable electronic device; CRT: cardiac resynchronisation therapy; ECG: electrocardiogram; EMI: electromagnetic interference; EOL: end-of-life; EP: electrophysiology; ERI: elective replacement indicator; ICD: implantable cardioverter-defibrillator; PVC: premature ventricular contractions
Initial clinical assessment
The first step in evaluating a patient with a suspected CIED problem is to determine whether the presenting symptoms are compatible with device malfunction. The timing of symptom onset is particularly informative. Sudden deterioration occurring immediately after implantation may suggest acute lead dislodgement, cardiac perforation, pocket haematoma, or pneumothorax [1]. In contrast, symptoms developing months or years after implantation are more likely to reflect lead failure, battery depletion, progressive myocardial disease, or the emergence of supraventricular arrhythmias [2].
Understanding the patient's underlying pacing indication is equally important. Patients implanted for intermittent sinus node dysfunction frequently tolerate transient pacing failure with only mild dizziness or fatigue. Conversely, individuals with complete atrioventricular block may rapidly develop profound bradycardia, recurrent syncope, or haemodynamic instability if ventricular pacing is lost. Likewise, patients with CRT devices may present primarily with worsening heart failure (HF) when effective biventricular pacing is not delivered, an occurrence usually associated with tachyarrhythmias [2].
Symptoms and signs that may suggest pacemaker malfunction, or could be device-related, include:
- recurrent syncope
- presyncope
- unexplained dizziness
- new/worsened exercise intolerance
- worsening HF symptoms
- persistent hiccups
- pectoral muscle stimulation
- device pocket pain, swelling, discharge, fever of unknown origin
Patients with ICDs require additional questioning regarding possible therapies delivered by the device. A single appropriate shock during ventricular tachyarrhythmia usually indicates normal device function rather than malfunction. In contrast, multiple shocks occurring without preceding symptoms, during normal activity or at rest, raise suspicion for oversensing, lead fracture, electromagnetic interference, or inappropriate arrhythmia detection. This distinction should guide the urgency of EP referral. ICD recipients may also describe audible beeps, tones, or vibrations of the device which are in fact patient notifier alerts indicating that the device has detected a condition requiring assessment (battery reaching the elective replacement indicator [ERI] stage, lead integrity problem, prolonged capacitor charging time, etc.), and therefore device interrogation [3,4].
The patient’s medication history provides essential information: anti-arrhythmic drugs or electrolyte-altering medications may increase pacing thresholds or exacerbate intrinsic conduction abnormalities, mimicking device malfunction. Similarly, recent magnetic resonance imaging, radiotherapy, thoracic surgery, or central venous catheter placement may provide important clues to mechanical lead injury or device reprogramming [2].
Finally, patients should be asked specifically about recent trauma involving the ipsilateral shoulder, vigorous upper limb movements, or repeated manipulation of the pulse generator. Such events may precede lead fracture or dislodgement.
Physical examination
Although physical examination rarely establishes a definitive diagnosis, it frequently identifies complications requiring urgent intervention before device interrogation.
Assessment begins with haemodynamic status. Hypotension, altered mental status, pulmonary oedema, or recurrent syncope in a patient with a pacing dependency should immediately raise concern for failure of ventricular pacing or loss of effective CRT. Such patients require urgent ECG assessment.
Pulse examinations also provide valuable information. Regular bradycardia below the programmed lower pacing rate suggests failure to pace or failure to capture. Conversely, irregular pulse deficits in CIED recipients may reflect frequent ventricular ectopy, atrial fibrillation, or inappropriate sensing.
Inspection of the device pocket should not be overlooked. Local erythema, tenderness, warmth, fluctuance, or purulent discharge strongly suggests CIED infection, while skin thinning or visible generator erosion represents a medical emergency requiring complete system extraction rather than conservative antibiotic treatment. Pocket swelling associated with skin bruising occurring shortly after implantation is usually caused by haematoma [5,6].
Inspection may also reveal obvious generator rotation. In these patients, careful questioning often reveals habitual manipulation of the device by the patient.
Observation for involuntary contractions of the pectoralis muscle, diaphragm, or upper abdominal wall may provide important clues to lead displacement, insulation failure, or cardiac perforation. Phrenic nerve stimulation may occur in CRT due to the close proximity of the left ventricular lead to the phrenic nerve. Patients describe rhythmic abdominal twitching or hiccup-like contractions and usually can be managed with device reprogramming.
Finally, signs of new/worsening HF may indicate pacing-induced cardiomyopathy or loss of effective ventricular synchrony in CRT recipients. In such cases, device interrogation is necessary to determine whether reduced biventricular pacing, atrial arrhythmias, or lead dysfunction is responsible.
ECG analysis
Following history and physical examination, the 12-lead ECG becomes the single most informative bedside investigation in suspected CIED malfunction. In many cases, careful analysis of pacing artefacts, sensing behaviour and QRS morphology allows the clinician not only to recognise abnormal device function but also to identify the underlying mechanism before formal device interrogation.
ECG troubleshooting
While device interrogation confirms the diagnosis, the ECG frequently identifies the underlying mechanism before specialised evaluation is available. A systematic interpretation also prevents unnecessary referrals prompted by normal pacing algorithms that may appear abnormal to the untrained observer.
It is important to obtain information from the CIED ID card or operative report if possible, including device model, lead models, acute intraoperative pacing and sensing thresholds, impedance values, and device programming.
A systematic five-step ECG approach:
Step 1. Determine the underlying rhythm
Step 2. Look for pacing artefacts
Step 3. Does every pacing spike produce depolarisation?
Step 4. Is intrinsic activity appropriately sensed?
Step 5. Compare the ECG with the clinical presentation
Figure 1.
Troubleshooting cardiac implantable electronic devices: practical ECG and chest x-ray examples.
Examples of CIED troubleshooting using 12-lead ECG and chest radiography. Panels 1-4 illustrate electrical pacing/sensing abnormalities; panel 5 shows normal fusion/pseudofusion; panels 6-9 illustrate mechanical lead/device complications.
a) Failure to pace (output failure)
Failure to pace refers to the absence of pacing stimuli despite an indication for pacing. On ECG, no pacing artefacts are observed even though the intrinsic rate falls below the programmed lower rate limit. This finding should immediately raise concern in pacing-dependent patients.
Typical ECG findings (Figure 1.1) [7]:
- bradycardia below programmed lower rate
- complete absence of pacing spikes
- long pauses
- escape rhythm or asystole in pacing-dependent patients
The differential diagnosis and possible causes include [2]:
- battery depletion beyond elective replacement interval
- pulse generator failure
- lead fracture with complete circuit interruption
- connector problems
- oversensing causing inappropriate inhibition
- inappropriate device programming
Symptoms depend almost entirely on pacing dependency. Patients with preserved intrinsic conduction may remain asymptomatic. Patients with complete atrioventricular block may present with:
- syncope
- recurrent presyncope
- hypotension
- acute HF
- cardiac arrest
Failure to pace in a pacing-dependent patient represents a medical emergency requiring immediate EP consultation. Conversely, an asymptomatic patient with intermittent sinus node dysfunction and preserved escape rhythm can usually undergo prompt – but not emergency – device interrogation.
Rarely, reversible conditions may transiently increase pacing thresholds and cause loss of capture. Potential causes include hyperkalaemia, acid–base disturbances, myocardial ischaemia, hypoxaemia, and certain antiarrhythmic drugs, particularly class I agents and amiodarone. These should be considered when pacing failure occurs without lead displacement or device malfunction.
b) Failure to capture
Failure to capture occurs when pacing output is delivered but fails to depolarise the myocardium. The ECG demonstrates a pacing spike without a subsequent P wave or QRS complex.
Atrial/ventricular failure to capture (Figure 1.2) [2]:
- atrial/ventricular pacing spike
- no immediate P wave/paced QRS
- escape ventricular rhythm may subsequently occur
CRT
- one ventricular lead captures while the other fails
- QRS morphology changes compared with previous ECGs in biventricular pacing or left bundle branch block (LBBB) pattern is present.
Capture failure may result from increased stimulation threshold or insufficient pacing output.
Common causes include [2]:
- lead displacement
- lead perforation
- exit block
- myocardial infarction involving the lead position
- severe electrolyte disturbances
- metabolic acidosis
- antiarrhythmic drugs
- progressive lead maturation
Acute loss of capture occurring within days after implantation strongly suggests lead dislodgement until proven otherwise. Months or years after implantation, lead fracture or rising capture thresholds become more likely. Symptoms are similar to failure to pace scenarios.
Whenever loss of capture is identified, a chest X-ray should be obtained to evaluate lead displacement. Nonetheless, every patient with documented loss of capture requires device interrogation.
c) Undersensing
Undersensing occurs when the device fails to recognise intrinsic electrical activity. Consequently, pacing continues despite spontaneous cardiac depolarisation. The device behaves as though intrinsic activity were absent.
Typical ECG findings (Figure 1.3):
- pacing immediately after spontaneous QRS complexes
- pacing during intrinsic P waves
- pacing within refractory periods
These pacing spikes may or may not capture the myocardium. If the myocardium is refractory, capture will not occur despite normal device function. This should not be mistaken for loss of capture.
Undersensing may result from [2]:
- lead dislodgement
- reduced intracardiac electrogram amplitude
- myocardial scar
- inappropriate sensing programming
Persistent ventricular undersensing warrants prompt device evaluation.
d) Oversensing
Oversensing is one of the most frequent causes of intermittent pacing failure. It occurs when the device detects electrical signals that should not be interpreted as cardiac depolarisation and is often intermittent. Because it falsely assumes intrinsic activity is present, pacing is inhibited.
ECG findings (Figure 1.4)
- unexpected pauses
- missing pacing spikes
- irregular inhibition of pacing
- intermittent profound bradycardia
Patients frequently report episodic dizziness rather than persistent symptoms. One of the most common sources of oversensing is lead fracture [1].
Clinical clues include:
- intermittent symptoms
- reproducible symptoms during arm movement
- normal ECG between episodes
Electrical signals generated by skeletal muscle activity may occasionally be detected by the device, resulting in myopotential oversensing. Symptoms often occur during weightlifting, vigorous arm movements, or isometric exercise. The resting ECG may appear completely normal.
Electromagnetic interference can lead to transient oversensing and may occur during electrocautery, industrial equipment exposure, or malfunctioning electrical installations. Symptoms disappear after removal of the interference source [3].
e) Fusion and pseudofusion
General cardiologists frequently refer patients because of fusion or pseudofusion beats which may be present despite completely normal device behaviour. ECG fusion beats appear when the paced and intrinsic impulses occur simultaneously, producing a QRS morphology between a paced and intrinsic complex (Figure 1.5). Pseudofusion occurs if a pacing spike is delivered immediately after intrinsic depolarisation has already begun; the pacing impulse does not contribute to depolarisation. No intervention is required. Recognition of these patterns avoids unnecessary referrals [1].
Patients receiving CRT deserve special attention. Even when pacing spikes are present, resynchronisation may be ineffective. The ECG should therefore be compared with previous paced tracings.
Warning signs include [2]:
- narrowing or widening of paced QRS
- disappearance of characteristic biventricular morphology
- return of native LBBB pattern
- atrial fibrillation with rapid ventricular response and intermittent ventricular pacing
Clinically, these patients often present with gradually worsening HF. Loss of effective CRT should prompt elective or urgent device interrogation depending on haemodynamic status.
f) Battery depletion
Modern generators provide months of warning before complete battery exhaustion. Although interrogation confirms battery status, several ECG clues may suggest battery depletion.
Examples include [1]:
- unexpected reduction in pacing rate to a manufacturer-specific backup rate
- loss of rate responsiveness (paced heart rate suddenly becomes fixed despite exercise)
- failure to pace (complete battery depletion)
Practical ECG red flags requiring EP referral
The following ECG findings should always prompt device interrogation:
- pacing spikes without myocardial capture
- absent pacing in a pacing-dependent patient
- pacing spikes occurring after intrinsic depolarisation (persistent undersensing)
- prolonged pauses suggestive of oversensing
- new change in paced QRS morphology suggesting lead displacement or CRT loss
- inappropriate ICD pacing or shocks
- unexplained fixed backup pacing rate suggesting battery depletion
Chest radiography
While the ECG evaluates electrical function, chest radiography provides information about the mechanical integrity of the pacing system. The two examinations are complementary: a normal ECG does not exclude structural abnormalities, and a normal chest X-ray does not rule out electrical lead dysfunction. Together, however, they identify most clinically relevant complications before device interrogation.
Standard posteroanterior and lateral chest radiographs should be obtained whenever device malfunction is suspected.
Assess the pulse generator
The examination begins with the pulse generator itself. Its location, orientation, and fixation should be evaluated before inspecting the leads. A generator that has rotated within the pocket may indicate Twiddler syndrome, Reel syndrome (Figure 1.6), or, less commonly, Ratchet syndrome. These syndromes result from repeated manipulation or gradual rotation of the device, causing the leads to retract from the heart [8].
Radiographic clues include:
- rotation of the generator compared with previous imaging
- coiling of one or more leads around the pulse generator
- shortening of the intracardiac lead course
- loss of normal lead slack
Patients often present with recurrent syncope, diaphragmatic stimulation, loss of pacing, or inappropriate ICD therapies. Referral to an electrophysiologist is warranted, as surgical revision is almost always required.
Evaluate lead position
Each lead should then be followed from the connector pin to its distal electrode.
Normal lead positions include:
- right atrial appendage or atrial septum for atrial leads
- right ventricular apex or interventricular septum for ventricular leads
- coronary sinus tributary for the left ventricular lead in CRT systems
Small variations in lead configuration are common and depend on implantation technique. Therefore, comparison with previous radiographs is more informative than relying solely on textbook appearances [9].
Lead dislodgement
Lead dislodgement remains one of the most common mechanical complications following CIED implantation, occurring most frequently within the first six weeks after the procedure.
Atrial lead dislodgement
Displacement of the atrial lead typically results in migration from the right atrial appendage towards the superior vena cava or right atrium.
The ECG frequently demonstrates:
- atrial loss of capture
- atrial undersensing
Ventricular pacing is not related to atrial activity. Chest radiography confirms migration of the lead tip from its original position (Figure 1.7).
Right ventricular lead dislodgement
Displacement of the ventricular lead is usually more clinically significant, particularly in pacing-dependent patients.
The ECG commonly demonstrates:
- ventricular loss of capture
- intermittent ventricular pacing
- ventricular undersensing
On chest radiography, the lead tip no longer occupies its original apical or septal position (Figure 1.8). Any patient with suspected ventricular lead displacement should undergo urgent device interrogation.
Left ventricular lead dislodgement
In CRT recipients, displacement of the coronary sinus lead often presents differently.
Patients usually develop:
- progressive dyspnoea
- worsening HF
- reduced exercise tolerance
Chest radiography usually reveals the withdrawal of the coronary sinus lead from its target branch (Figure 1.9).
Because loss of CRT may substantially worsen HF outcomes, these patients should be referred promptly, although emergency intervention is rarely required in haemodynamically stable individuals.
Lead fracture
Lead fracture usually develops years after implantation as a consequence of repetitive mechanical stress. The most common sites include beneath the clavicle (subclavian crush syndrome) and the costoclavicular space, immediately proximal to the pulse generator [2].
Radiographic findings may include:
- complete interruption of the conductor
- abrupt angulation ("kinking")
- focal narrowing
- discontinuity of the lead body
However, many fractures remain radiographically occult, particularly early conductor fractures. Therefore, a normal chest X-ray does not exclude lead fracture.
Clinical clues include:
- intermittent symptoms
- reproducibility during shoulder movement
- multiple ICD shocks
Whenever lead fracture is clinically suspected, device interrogation is mandatory regardless of radiographic findings.
Insulation failure
Unlike conductor fracture, insulation defects are rarely visible on chest radiography. Occasionally, obvious externalisation or abnormal lead contour may be observed, particularly in certain older lead models, but most insulation failures are electrical diagnoses established by interrogation.
Insulation failure often presents with:
- oversensing or intermittent loss of capture on ECG
- intermittent loss of capture
- multiple ICD shocks
Accordingly, a normal chest X-ray should never reassure the clinician when ECG findings strongly suggest electrical lead dysfunction.
Cardiac perforation
Although uncommon, lead perforation is an important complication that should be recognised promptly.
Patients may present with:
- pleuritic chest pain
- diaphragmatic pacing
- hiccups
- hypotension
- pericardial effusion
Chest radiography may demonstrate an abnormally distal lead tip extending beyond the expected cardiac silhouette. However, chest X-ray sensitivity is limited. If perforation is suspected clinically, transthoracic echocardiography—and often cardiac computed tomography—should be performed without delay [10].
Usefulness of remote monitoring
In clinically stable patients with non-urgent device findings, remote monitoring may be used instead of immediate EP referral. It can assess battery status, lead function, sensing, pacing parameters, arrhythmias, and stored electrograms, helping determine whether in-person evaluation is necessary. Symptomatic patients (syncope or clinically significant symptoms), those with suspected loss of capture, major lead abnormalities, or recurrent ICD therapies require prompt specialist assessment.
When should the general cardiologist refer the patient?
Clinical scenarios requiring either immediate electrophysiology referral or elective device interrogation are summarised below (Table 1).
Table 1. When should the general cardiologist refer the patient?
|
Immediate EP referral |
Elective device interrogation |
|---|---|
|
|
CRT: cardiac resynchronisation therapy; EP: electrophysiology; HF: heart failure; ICD: implantable cardioverter-defibrillator
Speaking to the patient
Patients presenting with suspected device malfunction are frequently anxious, particularly if they have experienced syncope or an ICD shock. Clear communication is essential.
Explain that most device-related symptoms can be evaluated using the ECG and chest X-rays before specialised testing is required. Reassure the patient that many apparent abnormalities represent normal device behaviour rather than a malfunction. Explain the symptoms that require immediate medical attention.
Providing patients with clear expectations regarding when to seek urgent medical care may reduce unnecessary emergency department visits while facilitating early recognition of clinically significant signs.
Conclusions
Although device interrogation remains the definitive diagnostic tool for evaluating CIED performance, it should not be considered the starting point of clinical assessment. Careful history taking, targeted physical examination, systematic ECG interpretation, and thoughtful analysis of chest radiography provide sufficient information to identify most clinically important abnormalities at the bedside.
For the general cardiologist, recognising the distinction between electrical malfunction, mechanical complications, and expected device behaviour is more valuable than understanding every programming parameter. Equally important is recognising when the device is not responsible for the patient's symptoms.
By adopting a structured approach, clinicians can confidently identify patients requiring urgent EP intervention, appropriately prioritise elective referrals, and safely reassure those with normally functioning devices.